Automatic Pensky Martene Flash Point Apparatus (ASTM D 93)
Accurate. Automatic. Compliant. Measure Safety Ensure Quality.
Flash Point is the lowest temperature at which application of the ignition source causes the vapour of a specimen to ignite under the specified condition of the test. The temperature of the approximately 70 ml test specimen is increased rapidly at first and then at a slower constant rate as the flash point is approached. At specified intervals, a test flame is passed across the cup, to measure the tendency of the test specimen to form a flammable mixture with air under controlled laboratory conditions.
Our goal is to deliver precision laboratory instruments and petroleum testing solutions that are reliable, automated, accurate, and built to provide long-term performance.
Standards
Conforms to international testing specifications and quality standards.
- Conforms to ASTM D93; IP 34; ISO 2719
Key Features & Benefits
Engineered with precise testing, automatic operation, touch screen control, and built-in storage memory.
- Range: 40 to 400° C
- Test Method: ISO 2719, ASTM D 93, IP 34, IS 1448 (P-21)
- Ignition Source: Electrical ignitor with LPG gas
- Temperature Sensor: PT-100 in stainless steel
- Flash Detector: CRC
- Controller: Touch Screen, Colour Display with PLC with audio visual alarm, 3-wire RTD, free potential flame detector with automatic barometric reading
- Inbuilt storage memory for Flash
- Flame-Sampling Arrangement: 10 RPM, D.C. 12 V or 24 V DC gear motor, vertical movement reciprocating to & fro
- Heater: 800 W, 230 V
- Electrical ignition flame-sampling with ignition transformer (continuous output 6 kV DC from 230 V AC input)
- Single Action Setting: Sensor and temperature probe mounted on a lid for easy setting
- Easy Operation: Select test mode, enter expected flash point, run automatic test, buzzer indication and result display on HMI
- Rugged, reliable and user-friendly design
Technical Specifications
| Temperature Range | 40 to 400° C |
|---|---|
| Test Methods | ISO 2719, ASTM D 93, IP 34, IS 1448 (P-21) |
| Ignition Source | Electrical ignitor with LPG gas |
| Temperature Sensor | PT-100 in stainless steel |
| Flash Detector | CRC |
| Controller | Touch Screen, Colour Display with PLC |
| Stirrer / Flame-Sampling | 10 RPM, D.C. 12 V or 24 V DC gear motor |
| Heater | 800 W, 230 V |
| Overall Size (L x W x H) | 280 mm (W) x 530 mm (D) x 415 mm (H) |
| Net Weight | 22 kg (49 lbs) |
| Electrical Requirement | 220-240V, 50/60Hz, Single Phase, 9A OR 115V, 50/60Hz, Single Phase, 8A |
| Dimension (in.) | 11 x 22 x 17 (28 x 53 x 41) |
| Shipping Weight | 78 lbs (35 kg) |
| Shipping Dimension | 4.7 Cu. ft. |
Safety Interlocks
Automatically shuts machine & horns buzzer in case of:
- EFP + 30°C or 400°C reached
- Temperature sensor failure
- Flash detector failure
- LPG solenoid malfunction
- Stirrer not working
- Heater failure
- Side ignition coil/solenoid failure
Utility
- Power Supply: 230 V AC, 16 A (50/60 Hz, Single Phase)
- Gas Supply: LP Gas or Natural Gas (Pressure < 8 Kpa) in case of gas fired.
Accessories Supplied
- Pensky Martene Cup & Lid – 1 No
- Thermal Printer – 1 No
- Thermal Printer Roll – 1 No
- PC Interface Cable – 1 No
- Software – 1 No
- Thermal Fuse – 1 No
- Signal Cable – 1 No
Applications
Petroleum & Petrochemicals
Testing fuels and petroleum products.
Lubricating Oils & Greases
Flash point evaluation for lubricants.
Fuels and Solvents
Safety and volatility testing.
Chemical & Pharmaceutical Industries
Industrial safety compliance testing.
Quality Control Laboratories
Routine quality assurance environments.
R&D and Academic Institutes
Research laboratories and educational testing.